It is with pride and gratitude that we reflect on the remarkable 10-year journey of European Journal of Arrhythmia & Electrophysiology. With the vital contributions of all of our esteemed authors, reviewers and editorial board members, the journal has served as a platform for groundbreaking research, clinical insights and news that have helped shape the […]

Cardiac Resynchronisation Therapy – An Approach to Difficult Left Ventricular Lead Placement

Mayank Singhal, Manoj K Rohit, Parag Barwad
Share
Facebook
X (formerly Twitter)
LinkedIn
Via Email
Mark CompleteCompleted
BookmarkBookmarked
Copy LinkLink Copied
Download as PDF
Published Online: Aug 20th 2015 European Journal of Arrhythmia & Electrophysiology, 2015;1(1):25–6 DOI: http://doi.org/10.17925/EJAE.2015.01.01.25
Select a Section…
1

Abstract

Overview

Left ventricular (LV) lead in cardiac resynchronisation therapy (CRT) is the most important and difficult lead to place, leading to abandonment of up to 10–15 % of procedures. Here we discuss various difficulties encountered in percutaneous placement of LV leads and what all can be done to ensure successful placement of the same and to prevent the already compromised patient from the requirement of epicardial lead placement.

Keywords

Cardiac resynchronisation therapy, left ventricular lead, percutaneous placement

2

Article

Of the three leads placed during cardiac resynchronisation therapy (CRT), the left ventricular (LV) lead is the most important and also needs to be placed precisely in the region that is activated last – hence it becomes the most difficult leads to be placed. With the advent of new and improved hardware and simultaneously more experience being gained by operators, percutaneous LV lead placement is now becoming more and more successful. However, up to 10–15 % of procedures are abandoned across the world due to failure of LV lead placement.1–2 Since the patients selected for CRT placement generally have a very low ejection fraction (EF) and can have deterioration in their clinical status because of slightest insult, one must look into the factors that can predict a difficult lead placement, so that necessary measures can be taken beforehand.

Failure to access the coronary sinus (CS) remains the most important reason for difficult LV lead placement.1–2 This may be due to enlarged right atrium (RA), severe tricuspid regurgitation, tortuous or vertically positioned CS ostium or because of a prominent Thebesian or Vieussens valve.3–4 A study has shown that a very high location of CS ostium (as determined by computed tomography [CT]) to be a sole predictor for prolonged CRT implantation procedures.5 Newly developed lead delivery sheaths with a large primary curve and a smaller secondary curve6 are helpful in cases of large RA, as these take support from lateral atrial wall and superior vena cava. A good method is to push the sheath into the right ventricle and then pull with an anticlockwise rotation. At times contrast injections may have to used to locate abnormally placed CS ostium. Deflectable mapping catheters (with or without central lumen) may be helpful in engaging CS ostium in difficult cases.

Another challenge that an operator might face is that of difficult anatomy of the coronary venous system including lack of suitable venous branches, sharply angulated or tortuous venous branches or branches with valves or stenosis. One may plan to elucidate the anatomy beforehand by venous phase of coronary angiogram in plain right anterior oblique or left anterior oblique views to eliminate any surprise during the procedure, or a balloon occlusive venogram may be taken once CS ostium has been cannulated. In cases of failure of such a technique, other methods including CT, intracardiac echocardiography and fibre optic endoscopy have been used,7–10 though data regarding the same are lacking. In case of a sharply angulated or tortuous target vein, one may use an appropriately shaped inner sheath that can selectively hook the desired vein and is capable of wire and lead delivery.6 Other techniques include pushing the wire as much inside the vein as possible to gain extra support, or to pull the wire while advancing the lead, or using second stiffer wire to reduce the tortuosity and provide extra support.11 Sometimes a different vein altogether may have to be chosen. Case reports of using stents to open up the vein or to stabilise the LV lead are also available. We presented a case at EuroPCR 2015, in which we had to place a coronary stent besides the precisely placed LV lead to stabilise it and to prevent distal migration, which was causing phrenic nerve stimulation and proximal placement causing prolapse into the RA. Another technique is to use retained guidewires for anchoring the LV lead. But all of these techniques make extraction of the lead near impossible if deemed necessary in future. Placement of the stent besides the lead may cause injury to the lead insulation.

Sometimes complications such as CS dissection may make visualisation of target vessels difficult, which may lead to a deferment of the procedure. In an interesting case report,12 such a case was treated with prolonged inflation of a coronary angioplasty balloon, which allowed completion of the procedure in the same attempt.

Despite all these factors sometimes the procedure has to be abandoned and analternative way of LV lead placement chosen. These include minimally invasive surgical alternatives (minithoracotomy, videoassisted thoracoscopic surgery and robotically assisted placement of LV leads), minimally invasive subxiphoid epicardial approach, transseptal endocardial left ventricular lead implantation and bifocal right ventricular pacing.13–18 In a prospective study19 comparing a percutaneous transvenous approach via the CS versus epimyocardial placement via a left lateral mini-thoracotomy, both epicardial and transvenous LV-lead placement for CRT therapy were found to be safe and effective. The transvenous group had a shorter intensive care unit stay (0.66 versus 3.8 days) and shorter ventilation times (0.34 versus 3.2 hours). At 6 months follow-up, no major differences in LV-lead parameters (threshold, sensing and impedance) were observed. In another study16 to assess the feasibility of transseptal endocardial LV pacing in patients in whom transvenous CS lead placement had failed, 10 such patients were taken up for endocardial LV lead placement, nine of whom could have the procedure completed successfully. The stimulation threshold was 0.78±0.24 V, and the R-wave amplitude was 14.2±9.7 mV. At 2 months follow-up, the stimulation threshold was 1.48±0.35 V with a 0.064±0.027 ms pulse width. There was no phrenic nerve stimulation observed in any of the patients. There were no thromboembolic complications at follow-up. LV transseptal endocardial lead implantation from the pectoral area was considered to be a feasible approach in patients with a failed CS approach and in whom epicardial surgical lead placement is not an option. Longer follow-up is warranted to determine the risk of thromboembolic complications.

To conclude, percutaneous LV lead placement though at times may be difficult, but it is still the preferred option. All measures should be taken to place the LV lead percutaneously as these patients have a severely compromised LV function and thus an additional burden of general anaesthesia and epicardial LV lead placement surgically can be detrimental.

2

References

1. Bax JJ, Abraham T, Barold SS, et al., Cardiac resynchronization
therapy: Part 2-issues during and after device implantation and
unresolved questions, J Am Coll Cardiol, 2005;46:2168–82.
2. Shepard RK, Ellenbogen KA, Challenges and solutions for
difficult implantations of CRT devices: the role of new
technology and techniques, J Cardiovasc Electrophysiol,
2007;18:21–5.
3. Gerber TC, Sheedy PF, Bell MR, et al., Evaluation of the
coronary venous system using electron beam computed
tomography, Int J Cardiovasc Imaging, 2001;17:65–75.
4. Shinbane JS, Girsky MJ, Mao S, Budoff MJ, Thebesian valve
imaging with electron beam CT angiography: implications
for resynchronization therapy, Pacing Clin Electrophysiol,
2004;27:1566–7.
5. Da Costa A, Gate-Martinet A, Rouffiange P et al. Anatomical
factors involved in difficult cardiac resynchronization therapy
procedure: a non-invasive study using dual-source 64-multislice
computed tomography, Europace, 2012;14:833–40.
6. Leon AR, New tools for the effective delivery of cardiac
resynchronization therapy, J Cardiovasc Electrophysiol,
2005;16:42–7.
7. Vaseghi M, Cesario DA, Ji S, et al., Beyond coronary sinus
angiography: the value of coronary arteriography and
identification of the pericardiophrenic vein during left
ventricular lead placement, Pacing Clin Electrophysiol,
2005;28:185–90.
8. Jongbloed MR, Lamb HJ, Bax JJ, et al., Noninvasive
visualization of the cardiac venous system using multislice
computed tomography, J Am Coll Cardiol, 2005;45:749–53.
9. Anh DJ, Chen HA, Eversull CS, et al., Early human experience
with use of a deflectable fiberoptic endocardial visualization
catheter to facilitate coronary sinus cannulation, Heart
Rhythm, 2006;3:875–8.
10. Shalaby AA, Utilization of intracardiac echocardiography to
access the coronary sinus for left ventricular lead placement,
Pacing Clin Electrophysiol, 2005;28:493–7.
11. Chierchia GB, Geelen P, Rivero-Ayerza M, Brugada P, Double
wire technique to catheterize sharply angulated coronary
sinus branches in cardiac resynchronization therapy, Pacing
Clin Electrophysiol, 2005;28:168–70.
12. Bosa F, Bethencourt M, Vargas M, et al., Prolonged inflation of
coronary angioplasty balloon as treatment for subocclusive
dissection of the coronary sinus during implantation of a
coronary sinus pacing lead, J Interv Card Electrophysiol,
2008;23:139–41.
13. Navia JL, Atik FA, Grimm RA, et al., Minimally invasive left
ventricular epicardial lead placement: surgical techniques
for heart failure resynchronization therapy, Ann Thorac Surg,
2005;79:1536–44.
14. Zenati MA, Bonanomi G, Chin AK, Schwartzman D,
Left heart pacing lead implantation using subxiphoid
videopericardioscopy, J Cardiovasc Electrophysiol,
2003;14:949–53.
15. Ji S, Cesario DA, Swerdlow CD, Shivkumar K, Left ventricular
endocardial lead placement using a modified transseptal
approach, J Cardiovasc Electrophysiol, 2004;15:234–6.
16. Van Gelder BM, Scheffer MG, et al., Transseptal endocardial
left ventricular pacing: an alternative technique for coronary
sinus lead placement in cardiac resynchronization therapy,
Heart Rhythm, 2007;4:454–60.
17. O’Donnell D, Nadurata V, Hamer A, et al., Bifocal right
ventricular cardiac resynchronization therapies in patients
with unsuccessful percutaneous lateral left ventricular
venous access, Pacing Clin Electrophysiol, 2005;28:27–30.
18. Vlay SC, Kort S, Biventricular pacing using dual-site right
ventricular stimulation: is it placebo effect?, Pacing Clin
Electrophysiol, 2006;29:779–83.
19. Doll N, Piorkowski C, Czesla M, et al., Epicardial versus
transvenous left ventricular lead placement in patients
receiving cardiac resynchronization therapy: results from
a randomized prospective study, Thorac Cardiovasc Surg,
2008;56:256–61.

3

Article Information

Disclosure

Parag Barwad, Manoj K Rohit and Mayank Singhal have no conflicts of interest to declare. No funding was received in the publication of this article.

Correspondence

Parag Barwad, Department of Cardiology, Advance Cardiac Center (ACC), PGIMER, Chandigarh, India. 160012, E: paragaims@gmail.com

Access

This article is published under the Creative Commons Attribution Noncommercial License, which permits any non-commercial use, distribution, adaptation
and reproduction provided the original authors and source are given appropriate credit