Pacemaker (PM) insertion is one of the most common invasive procedures in cardiology. This routine procedure may be associated with unfortunate complications such as myocardial perforation. Although this is a welldescribed but rare complication in symptomatic patients, subclinical perforation is even rarer and the diagnosis remains a challenge. In subclinical and late manifestation, it is even more difficult to diagnose.1
Conventional imaging modalities such as echocardiography and chest X-ray are helpful for diagnosis but may be limited by several factors. We present a case of myocardial perforation post-PM insertion with the lead migrating to the lateral part of the seventh rib, in which conventional imaging modalities such as 2D-echo or chest X-ray were unable to clarify the diagnosis.
Case presentation
A 94-year-old lady referred for recurrent syncope complicated by T10 spinal fracture and documented transient complete heart block with a 13 second pause while on holter monitoring, underwent single chamber pacemaker insertion (Medtronic active fixation 5776-58 cm lead). Two weeks later, she was seen for routine follow-up at our device clinic. The device interrogation revealed no sensing or pacing despite programming to the highest sensitivity and output and 4,000 Ohms for impedance. The patient was in sinus rhythm and experienced no recurrent syncope or any further cardiac symptoms.
Chest X-ray (Figure 1) was unable to determine whether the tip of the lead was extracardiac. 2D-echo did not add any additional information and ruled out pericardial effusion due to poor window in an obese
patient. A chest computed tomography (CT) scan showed the tip of the PM had perforated the right ventricle, extended 5.3 cm from epicardial fat and terminated along the costochondral junction of left seventh rib. This could be seen even better with the help of 3D pictures (Figures 2 and 3). The patient underwent an uncomplicated lead removal and a new passive ventricular lead insertion was performed with an R wave of 15.5 mV and threshold of 0.25 mV, and resistance of 921 Ohms. The patient was followed for three months post-operatively, with no changes
in measurements documented. This case highlights the usefulness of chest CT scan for the diagnosis of asymptomatic and subclinical ventricular perforation post-PM insertion in a patient presenting at the device clinic, in which lack of pacing at maximum output, loss of sensing and severely increased lead impedance were documented. Interestingly, this case did not show any pericardial effusion, as this is commonly seen with ventricular rupture.2 Even after manipulation and pulling back the PM lead, no pericardial tamponade occurred.3
Discussion
Although right ventricular (RV) perforation due to PM leads insertion is a rare complication, it could be potentially life threatening. Previously, it was thought that this happens more with active fixation leads,
however, in one recent study published by Migliore et al.4 this was not proved. This study showed no statistically significant difference in RV perforation between active versus passive fixation leads (0.5 versus 0.3%; p=0.3). In addition, there was no significant difference in RV perforation between PM and defibrillator leads.
Our case was done with an active fixation lead and was asymptomatic until PM data measurement post-implant. To explant and reposition the leads, we recommend having a cardiac surgeon on standby.