This website is intended for healthcare professionals only

Trending Topic

16 mins

Trending Topic

Developed by Touch
Mark CompleteCompleted
BookmarkBookmarked

Hypertension is the leading modifiable risk factor for global cardiovascular disease, responsible for an estimated 10.8 million deaths and more than 200 million disability-adjusted life years annually.1 Despite the availability of effective pharmacological and lifestyle interventions, prevalence continues to rise, particularly in low- and middle-income countries (LMICs), where over three-quarters of all cases now occur.2 The condition’s […]

53/Improving understanding of lesion parametrics in high-power short-duration ablation

G Panchal (Presenting Author) – Glenfield hospital, Leicester; A Mavilakandy – Glenfield hospital, Leicester; I Koev – Glenfield hospital, Leicester; A Kotb – Glenfield hospital, Leicester; M Ibrahim – Glenfield hospital, Leicester; M Lazdam – Glenfield hospital, Leicester; SH Chin – Glenfield hospital, Leicester; A Sandilands – Glenfield hospital, Leicester; R Somani – Glenfield hospital, Leicester; GA Ng – Glenfield hospital, Leicester
2 mins
Share
Facebook
X (formerly Twitter)
LinkedIn
Via Email
Mark CompleteCompleted
BookmarkBookmarked
Copy LinkLink Copied
Published Online: Oct 9th 2012 European Journal of Arrhythmia & Electrophysiology. 2022;8(Suppl. 1):abstr53
Select a Section…
1

Article

Background: Pulmonary vein isolation (PVI) is the gold standard management for patients with atrial fibrillation (AF) refractory to medical therapy. Contact-force (CF) sensing very-high-power short-duration (vHPSD) radiofrequency (RF) ablation (90 W/4 seconds) has emerged as a novel ablation modality. Previous RF ablation utilized Ablation Index (AI) as a surrogate marker of lesion quality to guide the operator, which is not utilized in QMODE+ (vHPSD) and thus, there is limited guidance on effective lesion formation. In this study, we investigated the correlation between certain lesion parameters to identify relationships and potential determinants for effective lesion formation. Furthermore, we studied the relationship between different anatomical locations on these parameters to gain further insight.

Method: A total of 60 consecutive AF patients (42 males, age 61.5 ± 9.06 years, 63% paroxysmal AF) underwent first-time PVI using QMODE+. All wide antral circumferential ablation (WACA) QMODE+ lesions (n=6040) were analysed for force-time integral (FTI), impedance drop, average CF, maximum temperature attained, and anatomical location. The anatomical regions assessed across both left and right WACA were posterior-superior (region 1-2), posterior-inferior (region 3-4), anterior-inferior (region 5-6), and anterior-superior (region 7-8), and pulmonary vein carina (region 9). All pulmonary veins (PVs) were subsequently checked with pacing manoeuvres to examine for gaps in ablation lesions and adenosine for acute pulmonary vein reconnection post-ablation.

Results: PVI was successful in all patients while first-pass isolation was observed in 30 patients (50%). A total of 6,040 lesions were performed, with 2,968 and 3,072 lesions in the left and right WACA, respectively. The average CF exhibited a positive correlation with maximum temperature attained and impedance drop (p<0.0001) while displaying a negative correlation with FTI (p<0.0001) (Figure 1). The PV carina (region 9) of the right WACA had the smallest impedance drop (p<0.0001) while the PV carina of the left WACA exhibited the lowest average CF and temperature attained (p<0.0001) (Figure 2). The highest number of gaps or acute reconnection (17 out of 30 patients) were seen in the region of the left pulmonary vein carina (p<0.0001).

Conclusion: To our knowledge, this is the first study that has investigated the characteristics of vHPSD ablation lesions according to different anatomical regions in the left atrium. CF was positively correlated with maximum temperature and impedance drop but negatively correlated with FTI. This is very unique to temperature-controlled ablation to reduce complications. Furthermore, ablation at both right and left PV carina demonstrated lower impedance drop and average CF which may explain difficult catheter positioning as well as thicker tissue. This study elucidates the relationship between the anatomical region, contact force, temperature and impedance drop, and will facilitate optimization for effective vHPSD lesion formation.