Trending Topic

3 mins

Trending Topic

Developed by Touch
Mark CompleteCompleted
BookmarkBookmarked

It is with pride and gratitude that we reflect on the remarkable 10-year journey of European Journal of Arrhythmia & Electrophysiology. With the vital contributions of all of our esteemed authors, reviewers and editorial board members, the journal has served as a platform for groundbreaking research, clinical insights and news that have helped shape the […]

19/The impact of generic ICD programming on clinical outcomes

P Khan (Presenting Author) - King’s College Hospital NHS Trust, London; N Jahagirdar - King’s College Hospital NHS Trust, London; M Laybourn - King’s College Hospital NHS Trust, London; I Harding - King’s College Hospital, London; A Cannata - King’s College Hospital NHS Trust, London; D Bromage - King’s College Hospital, London; F Murgatroyd - King’s College Hospital, London; PA Scott - King’s College Hospital, London
Share
Facebook
X (formerly Twitter)
LinkedIn
Via Email
Mark CompleteCompleted
BookmarkBookmarked
Copy LinkLink Copied
Published Online: Oct 3rd 2011 European Journal of Arrhythmia & Electrophysiology. 2021;7(Suppl. 1):abstr19
Select a Section…
1

Article

Background: Inappropriate implantable cardioverter defibrillator (ICD) therapy can significantly reduce quality of life and is associated with adverse outcomes. Introduction of generic programming strategies, combing high detection rates and long detection times have shown to significantly reduce inappropriate therapy. Each of the 3 large randomised controlled trials that have informed programming strategies included only a single device manufacturer, with tightly controlled programming parameters in both the active and control groups. Both detection time and lower treatment zone rate varied and no study evaluated combining the most aggressive programming of each parameter. A standardised programming protocol was introduced at our centre in October 2013 with shock reduction programming extrapolated from trials of one manufacturer and applied to across manufacturers.

Objective: In 2017 a series from 4 institutions reported 10 cases where normally functioning ICDs failed to deliver timely therapy for ventricular fibrillation (VF). The authors suggested that complex and unanticipated interactions between manufacturer-specific features and generic programming prevented therapy for VF. The purpose of this retrospective cohort study was to evaluate the evaluate the benefits and safety of generic programming and explore the effect of different detection programming combinations on outcomes.

Methods: We included all new ICDs in a single centre (2009–2019). In 2013 a standardised programming protocol based on generic programming was introduced, incorporating high detection rates (200 bpm for primary prevention) and long detection (30/40 or equivalent in VF zone) for all patients. Patients were classified into three groups based on implant programming: pre-guideline (PS), post-guideline and guideline complaint (GC), and post-guideline but not guideline complaint (NGC). The endpoints were the first occurrence of any device therapy (ATP or shocks) or shock alone, syncope and all-cause mortality. Survival analysis was used to evaluate outcomes. To evaluate the potential impact of combining long detection time and a high lower rate detection zone, we constructed a composite ‘programming’ score, which combined measures of both parameters. The relationship between this composite programming score and study endpoints were evaluated in separate multivariable Cox proportional hazards models, with separate analyses for primary and secondary prevention patients.

Results: 1,003 patients were included (mean follow-up 1,519 ± 1,005 days). In primary prevention patients (n=583) freedom from ICD therapy (91.5% vs. 73.6%; p<0.001) or shock (94.7% vs 84.8%; p=0.004) were significantly higher in GC compared to PS patients, without significant increase in syncope or mortality. In secondary prevention patients (n=420) freedom from any ICD therapy or any shock were non-significantly higher in GC compared to PS patients, without an increase in syncope or mortality. Multivariant analysis of composiste programming scores found a lower rate of 200 bpm with long detection provided the greatest reduction in risk of ICD therapy without any demonstrable increase in syncope or mortality.

Conclusion: In a real-world population, introduction of a standardised programming protocol significantly reduced the burden of ICD therapy without an increase in adverse outcomes. Furthermore, the use of generic programming appeared effective across manufacturers.